MRA Impact on Market Access

When CE Marking Is Sufficient

When MRA Chapter 4 is fully operational:

  • CE-marked devices compliant with EU MDR/IVDR can generally be placed on the Swiss market with simplified DAPI notification (not full registration) for Class I devices; Class IIa and above still require full DAPI registration even with MRA operational status
  • EU NB certificates are recognised by Swissmedic without separate technical review
  • Technical documentation is not separately re-evaluated by Swissmedic

When Separate Swiss Registration Is Required

When MRA Chapter 4 is not fully operational:

  • Full DAPI registration is required for Class IIa, IIb, III and IVD List A/B/Self-test
  • Class I devices: DAPI notification within 30 days (same as when MRA is operational)
  • EU MDR/IVDR NB certificates and DoC are used as the conformity evidence within the DAPI registration

Always Required — Regardless of MRA Status

ObligationMRA OperationalMRA Not Operational
CH REP appointment✓ Required✓ Required
DAPI notification/registration✓ Simplified✓ Full registration
Swiss labelling (DE/FR/IT)✓ Required✓ Required
Vigilance reporting to Swissmedic✓ Required✓ Required
FSCA notification to Swissmedic✓ Required✓ Required

Current MRA Status

caution

Always verify the current MRA Chapter 4 operational status before planning your Swiss market access timeline. See MRA Status Updates and contact Swissmedic or SECO directly.

Disclaimer

AI-assisted navigation aid only. Always verify against official Swissmedic and Fedlex sources. Not legal or regulatory advice.

MRA Chapter 4 operational status should be verified within 30 days of commencing Swiss market access planning, as status changes impact registration timeline and may require either simplified notification or full registration.