SaMD & MDSW Regulation

Draft MDSW Circular 2025 · CDRRHR · IEC 62304 · ASEAN AMDD

Scope — What Is MDSW Under the Philippines Framework?

The draft Philippines MDSW circular covers software intended to be used for a medical purpose — diagnosis, prevention, monitoring, treatment, or alleviation of disease or injury. Two categories apply:

CategoryDefinitionExample
SaMDSoftware that is itself a medical device, not embedded in hardwareAI diagnostic imaging analysis app, ECG interpretation software
SiMDSoftware embedded in or controlling a hardware medical deviceInfusion pump controller firmware, pacemaker programming software

Out of scope:

  • Software with no medical purpose (hospital administration, scheduling)
  • IVD software (separate framework pending)
  • Software used for manufacturing or quality management of devices

Classification of SaMD

SaMD classification uses ASEAN AMDD Class A–D, informed by two dimensions:

1. Significance of the information provided:

  • Drives treatment/diagnosis (highest significance)
  • Informs clinical management (moderate significance)
  • Informs lifestyle decisions / not life-critical (lower significance)

2. State of the healthcare situation:

  • Critical — life-threatening or irreversible harm possible if wrong
  • Serious — significant deterioration possible
  • Non-serious — appropriate management, no significant change expected

The intersection of these determines class. A SaMD that drives a treatment decision in a critical condition is likely Class D; one that informs a lifestyle decision in a non-serious condition may be Class A.

Technical Documentation for MDSW

In addition to the standard CSDT sections, MDSW submissions must include software-specific documentation:

Software Documentation Requirements

DocumentDescription
Software descriptionOverview of software functions, architecture, operating environment (OS, hardware dependencies)
Intended medical purposeClear statement of what the software does clinically and for whom
Software Development Lifecycle (SDLC)Alignment with IEC 62304 — software development, maintenance, problem resolution
Software requirements specificationFunctional and safety requirements
Software architectureDescription of software items and their interactions
V&V test documentationUnit testing, integration testing, system testing records
Known anomaliesList of known software defects and risk mitigation
Revision historyVersion history and change management records
Cybersecurity documentationThreat model, security controls, vulnerability management plan
Post-market update planHow software updates will be managed and submitted post-approval

IEC 62304 — Software Lifecycle Standard

IEC 62304 is the recognised international standard for medical device software lifecycle processes. Compliance with IEC 62304 is expected for MDSW submissions.

IEC 62304 defines three software safety classes:

  • Class A: No injury or damage to health possible
  • Class B: Non-serious injury possible
  • Class C: Death or serious injury possible

Requirements scale with the safety class — Class C requires the most rigorous development documentation.

Cybersecurity for Connected MDSW

Connected MDSW (software with network connectivity, data sharing, or remote update capability) must address cybersecurity risks:

Cybersecurity RequirementDetail
Threat modellingIdentify potential attack vectors and threat actors
Secure architectureDefence in depth, principle of least privilege
Authentication and access controlUser authentication, role-based access
Data integrityProtection against unauthorised modification of data
EncryptionEncryption of data in transit and at rest where appropriate
Vulnerability disclosureProcess for receiving and responding to vulnerability reports
Update managementSecure delivery and installation of software updates

Post-market cybersecurity monitoring must be part of the PMS plan.

Post-Approval Software Updates

Software updates after CMDN/CMDR approval need careful management:

  • Minor bug fixes with no change to intended purpose or safety profile: likely no variation required but document and maintain records
  • New features or expanded intended purpose: likely require a variation application before release
  • Safety-critical updates / patches: notify CDRRHR promptly and follow FSCA procedures if the update addresses a safety issue

The draft circular is expected to provide more specific criteria for which updates require regulatory notification. Until finalised, adopt a conservative approach and consult CDRRHR for significant updates.

Preparing Now

While the MDSW circular is pending finalisation, manufacturers can prepare:

  • Complete IEC 62304-aligned SDLC documentation
  • Prepare a comprehensive V&V test report
  • Conduct a cybersecurity threat modelling exercise
  • Draft a post-market update management procedure
  • Classify the software under the ASEAN AMDD Class A–D framework

These documents will be required regardless of the final circular text and can be adapted when requirements are confirmed.