In Vitro Diagnostics (IVD)
IVD Regulation in Malaysia
In vitro diagnostic medical devices (IVDs) are regulated under Act 737 and the Medical Device Regulations 2012, using a classification system based on GHTF SG5 IVD guidance with Annex II lists for high-risk IVDs.
This page covers IVD-specific obligations beyond the classification framework (see IVD Classification for classification detail).
Performance Evaluation — The Core of IVD Conformity
For IVDs, performance evaluation (comprising analytical performance and clinical performance studies) is the primary conformity assessment pathway and central evidence requirement for demonstrating safety and performance. It encompasses:
Analytical Performance
The ability of the IVD to correctly measure what it is intended to measure:
| Parameter | Description | Key Standard |
|---|---|---|
| Accuracy/Trueness | Closeness of test result to true value | ISO 5725, ISO 17511 |
| Precision | Repeatability (within-run) and reproducibility (between-run, between-day) | EP05 (CLSI) |
| Analytical sensitivity | Limit of Detection (LoD) and Limit of Quantitation (LoQ) | EP17 (CLSI) |
| Analytical specificity | Absence of interference from cross-reacting substances | EP07 (CLSI) |
| Linearity | Range over which the relationship between result and concentration is linear | EP06 (CLSI) |
| Measuring interval | Reportable range of the device | |
| Stability | Reagent, calibrator, and in-use stability |
Clinical Performance
The ability of the IVD to correctly identify patients with or without the target condition:
| Parameter | Description |
|---|---|
| Diagnostic sensitivity | True positive rate — proportion of diseased patients correctly identified as positive |
| Diagnostic specificity | True negative rate — proportion of non-diseased patients correctly identified as negative |
| Positive predictive value | Probability that a positive result is a true positive |
| Negative predictive value | Probability that a negative result is a true negative |
Clinical performance studies must use samples from clinically well-characterised patients and be compared against an appropriate reference method or clinical reference standard.
Metrological Traceability
For quantitative IVDs, MDA expects that calibrators are traceable to:
- International reference materials (where available — check JCTLM database)
- International reference measurement procedures
- SI units where applicable
Traceability documentation must be included in the technical file.
Reference Interval Studies
For IVDs providing results compared to reference intervals, manufacturers should:
- Provide reference intervals relevant to the Malaysian and/or Asian population where possible
- Reference published studies on Asian population reference intervals
- Justify the use of reference intervals derived from non-Asian populations if Malaysian data is unavailable
Reference intervals derived entirely from European or North American populations may not be appropriate for Malaysian patients.
Kit Components
For kit-based IVDs, the technical file must cover all components supplied:
- Reagents (all formulations)
- Calibrators
- Controls
- Diluents/buffers
- Ancillary materials
Each component's composition, function, and stability must be documented.
Instrument-Reagent Compatibility
For IVDs designed to work with a specific analyser or instrument platform:
- Validated combinations of reagent lots and instrument models must be specified
- Performance data must be generated on the specific instrument(s) intended for use
Post-Market Surveillance for IVDs
IVD-specific PMS activities include:
- Monitoring of external quality assurance (EQA) scheme results
- Tracking of complaints related to anomalous results
- Literature surveillance for emerging data on the analyte and clinical condition
- Monitoring of competitor recalls and safety alerts for similar devices
- Post-market performance follow-up studies for Annex II IVDs
Labelling for IVDs
IVD labels and IFUs must include (in addition to general labelling requirements):
- Intended analyte and matrix (specimen type)
- Principles of the procedure
- Analytical measurement range and units
- Limitations of the procedure
- Expected values / reference intervals
- Interferences identified
- Instructions for quality control
See Labelling Requirements for the complete list.
Complete the truncated section: 'Labelling for IVDs — In addition to general requirements, IVD labels and IFUs must include: Analyte and specimen type, Analytical measurement range, Reference intervals (where applicable), Interferences, Principles of the procedure, and a warning or statement if the IVD is for in vitro use only. See Labelling Requirements for full detail on mandatory label elements and language requirements for IVDs supplied in Malaysia.'