Digital Health Framework

NMRA · NMRA Act No. 5 of 2015 · SaMD · IEC 62304

Overview

Sri Lanka's NMRA regulates Software as a Medical Device (SaMD) under the same legal framework as physical devices — the NMRA Act No. 5 of 2015. Software that meets the medical device definition must be registered before being marketed or supplied in Sri Lanka.

Classification of SaMD follows the NMRA's EU-derived classification framework. See SaMD, Borderline & Combination Products for classification rules.

Regulatory Basis

InstrumentRelevance
NMRA Act No. 5 of 2015Medical device definition includes software
NMRA Classification GuidelinesApply EU-derived risk rules to SaMD
IEC 62304Expected software lifecycle standard in technical dossier
ISO 14971Risk management including software hazards
IMDRF SaMD N41International principles informing NMRA approach

What Qualifies as SaMD in Sri Lanka

Software is a medical device when its intended purpose is to diagnose, prevent, monitor, treat, or alleviate a disease, injury, or disability. Classification follows the general device classification framework with software-specific considerations for:

  • The severity of the disease or condition addressed
  • Whether the software drives a clinical decision or merely provides information
Software TypeMedical Device?
AI diagnostic imaging analysisYes
Clinical decision support driving treatmentYes
Remote patient monitoring with clinical alertsYes
Drug dosing calculator with clinical outputYes
General wellness / fitness trackingNo
Administrative EMR (no clinical output)No

Technical Documentation for SaMD

SaMD submissions in the F-MDR-035 consolidated dossier must include software-specific documentation in addition to standard requirements:

DocumentStandard
Software description and intended medical purpose
Software Development Lifecycle documentationIEC 62304
Software requirements and architecture
Verification and validation test reports
Usability engineeringIEC 62366-1
Cybersecurity risk assessmentIMDRF N60 / IEC 81001-5-1
Algorithm validation (AI/ML)IMDRF AI/ML guidance
Risk management including software hazardsISO 14971
Post-market update management plan

Registration Pathway for SaMD

SaMD follows the standard NMRA two-step registration pathway:

  1. Manufacturing site registration (for the software developer's facility)
  2. Device registration — full dossier or reliance pathway where applicable

For the reliance pathway, if the SaMD has been approved by a recognised Reference NRA (Australia, Canada, EU, Japan, US, UK, Switzerland), the NMRA may rely on that assessment. This is the fastest route for SaMD with existing major-market approvals.

AI and Machine Learning

AI/ML-based SaMD is regulated as a medical device where it meets the intended purpose criterion. Key considerations:

  • Training data: Document representativeness of training data for the intended patient population
  • Algorithm validation: Validate on independent datasets; provide performance metrics (sensitivity, specificity, AUC or equivalent)
  • Adaptive algorithms: Post-deployment learning that changes clinical output may create change notification obligations
  • Clinical evidence: High-risk AI tools require clinical validation demonstrating diagnostic accuracy

Cybersecurity

Connected SaMD (with network or cloud connectivity) must address cybersecurity risks:

  • Threat modelling — identify attack vectors
  • Secure architecture — authentication, access control, encryption
  • Vulnerability management — process for receiving and responding to security vulnerabilities
  • Software update mechanism — how security patches and feature updates are managed
  • Post-market cybersecurity monitoring plan

Post-Approval Software Updates

Changes to SaMD post-approval must be managed:

  • Safety-critical updates addressing a known risk: follow FSCA procedures — notify NMRA before releasing the update if it addresses a safety issue
  • Changes to intended purpose or significant new features: Submit a change notification or variation application before release
  • Minor bug fixes with no clinical or safety impact: Document and maintain records; no variation typically required

Engage NMRA for guidance on specific updates, as the boundary between notifiable and non-notifiable changes in SaMD is not yet codified in detailed NMRA guidance.